Loop impedance with 0.01 Ω resolution, without tripping RCDs

AMZC-306

Ohmmeters and loop impedance meters

Image: Sonel MZC-306 loop impedance meter with orange casing; the display shows 0.63 Ω, with keypad and rotary selector.

AMZC-306

Short-circuit loop impedance meter

C/ María Barbeito, 14Contact Contact T [+34] 982 20 99 20
27003, Lugo, SpainT [+34] 982 20 99 20 info@amperis.com | www.amperis.com

Short-circuit loop impedance with 0.01 Ω resolution and calculated fault current in networks up to 750 V.

Image: Front panel with function selector and display

The whole loop, in detail

The AMZC-306 measures the short-circuit loop impedance ZL-PE, ZL-N and ZL-L with 0.01 Ω resolution over 0.13 to 1999 Ω per IEC 61557, in installations from 110/190 V to 400/690 V (100 to 750 V range). It separates the resistive and reactive components, calculates the prospective short-circuit current for the nominal voltage and measures with high current up to 30 A.

No nuisance tripping

The ZL-PE RCD mode measures the loop without tripping residual-current devices rated above 30 mA in the 100-440 V range. The meter checks the correct connection of the PE terminal with a contact electrode, keeps measuring when L and N are swapped and warns of overvoltage, and it measures voltage up to 750 V with 0.1 V resolution below 250 V.

Main features

  • Loop impedance resolution of 0.01 Ω from 0.13 to 1999 Ω.
  • RCD mode without tripping devices rated above 30 mA.
  • Resistive and reactive components and calculated fault current.
  • PE terminal check by contact electrode; works with L and N swapped.
  • Memory of 990 cells (3500 entries) with USB connection to a PC.
  • More than 5000 measurements on one set of alkaline batteries.

Applications and delivery

Applications

  • Verification of protection against electric shock in LV networks.
  • Fault-loop measurements in industrial installations up to 690 V.
  • Circuits protected by residual-current devices.
  • Periodic inspections with documented results.

Ready to work

Rated CAT IV 600 V (CAT III 1000 V) with double insulation and IP54 protection, the AMZC-306 switches off automatically after 300 seconds and is supplied with the UNI-SCHUKO adapter, 1.2 m test leads, probes, crocodile clips, carrying case, harness, USB cable, NiMH battery pack and calibration certificate; 5, 10 and 20 m leads and three-phase socket adapters are optional.

The stored measurements are transferred to the PC, reviewed and turned into verification reports.

Software & applications

Image: Preview of a verification report of the installation, ready to print or export

From the memory to the PC

The free PC reading software downloads through the USB cable the results stored in the 990 memory cells of the AMZC-306 — up to 3500 entries — and shows them on screen ready to review. Every measurement keeps its description of client, site and measuring point, so complete installations are checked circuit by circuit without transcription errors. The software runs on any Windows computer.

Reports and export

Results are printed directly or exported to popular formats for spreadsheets and databases. The measured loop impedances and prospective short-circuit currents are gathered into clear tables that support the verification protocol of the installation, and the stored data builds a history of each circuit that reveals its evolution between inspections.

What the software provides

  • Free reading software for Windows.
  • Complete transfer of the memory through the USB port.
  • Direct printing and export to popular formats.
  • Data organised by client, site and measuring point.
AMZC-306 — TECHNICAL SPECIFICATIONS
Loop impedanceZL-PE, ZL-N, ZL-L; resolution 0.01 Ω
Measuring range0.13…1999 Ω (1.2 m leads); 0.25…1999 Ω (UNI-SCHUKO), IEC 61557
Accuracy±(5 % m.v. + 3 digits) to 19.99 Ω; ±(4 % m.v. + 3 digits) above
RCD modeNo tripping of RCDs >30 mA; 0.43…1999 Ω; 100-440 V
Networks110/190 V to 400/690 V; range 100…750 V; 45-65 Hz
Short-circuit currentCalculated for the nominal voltage; R and X components
High-current measurementUp to 30 A
Voltage measurement0…750 V; 0.1 V resolution up to 250 V
Memory and interface990 cells (3500 entries); USB
Power supply4 × LR14 alkaline or NiMH batteries; >5000 measurements
Auto-offAfter 300 seconds
SafetyCAT IV 600 V (CAT III 1000 V); double insulation
Protection classIP54, per EN 60529
Operating conditions0…+45 °C; humidity 20-80 %
Dimensions and weight288 × 223 × 75 mm; 2 kg
StandardsIEC 61010-1, IEC 61557, EN 60529

AMZC-306 — MEASUREMENT DATA

Voltage measurement

Display rangeResolutionAccuracy
0.0…249.9 V0.1 V±(2% m.v. + 4 digits)
250…750 V Frequency range: 45…65 Hz.1 V±(2% m.v. + 2 digits)
Frequency range: 45…65 Hz.

Fault loop impedance ZL-PE, ZL-N, ZL-L

Measuring range acc. to IEC 61557, from 0.13 Ω (1.2 m leads) / 0.17 Ω (5 m) / 0.21 Ω (10 m) / 0.29 Ω (20 m) / 0.19 Ω (UNI-Schuko adapter) up to 1999 Ω.

Fault loop impedance ZL-PE RCD (without tripping RCDs)

Measuring range acc. to IEC 61557, from 0.43 Ω (1.2 m leads) / 0.47 Ω (5 m) / 0.51 Ω (10 m) / 0.59 Ω (20 m) / 0.49 Ω (UNI-Schuko adapter) up to 1999 Ω.

Display rangeResolutionAccuracy
0…19.99 Ω0.01 Ω±(6% m.v. + 10 digits)
20.0…199.9 Ω0.1 Ω±(6% m.v. + 5 digits)
200…1999 Ω1 Ω±(6% m.v. + 5 digits)

Does not trip RCDs with IΔn ≥ 30 mA. Rated voltages Un: 110, 115, 127, 220, 230, 240, 390, 400 V; operating range 100…440 V. RS and XS for ZS < 200 Ω: ±(6% + 10 digits) of ZS up to 19.99 Ω, ±(6% + 5 digits) above.

Display rangeResolutionAccuracy
0…19.99 Ω0.01 Ω±(5% m.v. + 3 digits)
20.0…199.9 Ω0.1 Ω±(4% m.v. + 3 digits)
200…1999 Ω1 Ω±(4% m.v. + 3 digits)

Rated voltages Un: 110/190, 115/200, 127/220, 220/380, 230/400, 240/415, 290/500 and 400/690 V; operating range 100…440 V (ZL-PE, ZL-N) and 100…750 V (ZL-L); fn 50/60 Hz; maximum test current 36.7 A (10 ms) at 690 V. RS and XS indicated for ZS < 200 Ω: ±(5% + 5 digits) of the ZS value. "m.v." – measured value.

Indication of the prospective short-circuit current IK

Display rangeResolutionAccuracy
0.055…1.999 A0.001 Acalculated from the fault-loop accuracy
2.00…19.99 A0.01 Acalculated from the fault-loop accuracy
20.0…199.9 A0.1 Acalculated from the fault-loop accuracy
200…1999 A1 Acalculated from the fault-loop accuracy
2.00…19.99 kA0.01 kAcalculated from the fault-loop accuracy
20.0…69.0 kA0.1 kAcalculated from the fault-loop accuracy

Calculated from the unrounded fault-loop impedance ZS and the rated voltage (IEC 61557); indication up to 69.0 kA in the standard measurement and up to 40.0 kA in the RCD mode.