Accessible HTML version of the document: the same text as the PDF, with headings, tables and a description of each image. Download the original PDF (PDF · 3 pages · 173 KB)

Insulation Resistance Meter

HIGHEST ACCURACY & LOWEST COST

Image: Insulation resistance tester in an open grey rugged case, with orange-framed display, keypad, measurement terminals and rotary selector.

Insulation resistance measurement - 40TΩ: * 50 – 1000 V, 25 V steps * 1000 – 10000 V, 50 V steps Memory, bluetooth keyboard (option), burnout location. Capacitance and DC/AC voltage measurement. Leakage current measurement. Digital filters function. Measurement of protective connections and equipotential bonding.

AMPERIS PRODUCTS S.Lagricultura, 34, Contact 27003 Agricultura,34 +T [+34] 982 20 99 20 | F [+34] 982 20 99 11 lugo [spain]27003, Lugo, Spain info@amperis.com | www.amperis.com

Main characteristics are the following:

Insulation resistance measurement:

  • Measurement voltage: *50 – 1000 V, 25 V steps *1000 – 10000 V, 50V steps -Continuous indication of measured insulation resistance or leakage current. -Automatic discharge of measured object capacitance voltage. -Acoustic signaling of 5 seconds intervals to facilitate capturing time characteristics. -Metered T1 , T2 and T3 test times for measuring one or two absorption coefficients from the range of 1…600 s. -Adjustable measuring time to 99’59’’. -Polarization index (PI) and dielectric absorption ratio (DAR) measurement. -Indication of actual test voltage during measurement. -1.2 mA, 3mA and 5 mA test current. -Insulation resistance measurement using twoor three-wire method. -Measurements with test leads up to 20 m. -Protection against measuring live objects.-Measurement of capacitance during the measurement of RISO. -Measurement of temperature (with additional probe - WASONT1). -Step voltage insulation resistance measurement (SV). -Dielectric Discharge calculation (DD). -Location of damage (burnout). Digital filters function for measurements in high noise environment. Continuity measurement of protective connections and equipotential bonding in accordance with EN 61557-4 with current > 200 mA. Adjustable limits for measured resistance RISO and RCONT. Measurement of leakage current during insulation resistance testing. DC and AC voltage measurement in the range of 0…750 V. Drawing graphs on the display during measurement. Innovative memory with possibility of description of: measurement points, facilities,… Operating with mini Bluetooth keyboard (option). Graphic LCD 5,6’’ backlit. Power supply from main power line or battery packs. Built-in fast charger.

Measurement of capacitance

Range ResolutionAccuracy
1...999 nF 1 nF 1,00...49.99 μF 0,01 μF±(5%m.v. + 5 digits)
Capacity measurement result is displayed after the RISO measurement. For measuring voltages under 100 V capacitance measurement accuracy not specified.

Insulation resistance measurement

Measurement range acc. to EN 61557-2: 5 MΩ...20,0TΩ (U N =5000V)

RangeResolutionAccuracy
0,0...999,9kΩ1kΩ
1,000...9,999MΩ0,01MΩ
10,0...99,99MΩ0,1MΩ±(3%.m.v.. + 10 digits)
100,0...999,9MΩ1MΩ±(3%.m.v.. + 10 digits)
1,000...9,999GΩ0,01GΩ
10,0...99,99GΩ0,1GΩ
100,0...999,9GΩ1GΩ±(3,5%.m.v.. + 10 digits)
1,000...9,999TΩ0,01TΩ±(7,5%.m.v.. + 10 digits)
10,0...40,00TΩ0,1TΩ±(12,5%.m.v.. + 10 digits)

Values of measured resistance depending on measurement voltage

Voltage UISOMeasurement rangeMeasurement range (AUTOISO-5000)
50V200GΩ20GΩ
100V400GΩ40GΩ
250V1,00TΩ100GΩ
500V2,00TΩ200GΩ
1000V4,00TΩ400GΩ
2500V10,00TΩ400GΩ
5000V20,00TΩ400GΩ
10000V40,00TΩ-----

Continuity measurement of protective connections and equipotential bonding, I=200mA

RangeResolutionAccuracy
0.00...19.99 Ω0.01 Ω±(2%m.v. + 3 digits)
20.0...199.9 Ω0.1 Ω±(2%m.v. + 3 digits)
200...999 Ω1 Ω±(4%m.v. + 3 digits)
Voltage on open terminals: 4...24 V. Output current at R <15 Ω: min. 200 mA (I
Output current at R <15 Ω: min. 200 mA (IS Compensation of test lead resistance. Current flowing in both directions, mean vaSC : 200...250 mA). alue of resistance is displayed.

DC and AC voltage measurement

RangeResolutionAccuracy
0,0...29.9 V0,1 V±(2%m.v. + 20 digits)
30,0...299.9 V0,1 V±(2%m.v. + 6 digits)
300...750 V1 V±(2%m.v. + 2 digits)
Frequency range: 45...65Hz.

Measurement of temp

RangeResolutionAccuracy
-40,0...99,9 °C1 °C±(3%m.v. + 8 digits)
-40,0...211,8 °F1 °F±(3%m.v. + 16 digits)

“m.v.” = “measured value”.

AMPERIS PRODUCTS S.Lagricultura, 34, 27003 Agricultura,34 lugo [spain]27003, Lugo, Spain

Specifications AMIC-10k1

Electric security: Type of insulationDouble, acc. to EN 61010-1 and IEC 61557 Measurement categoryCAT IV 600V (III 1000V) EN 61010-1 Protection class acc. to EN 60529IP40 (IP67 with cover closed) Other technical data: Power supply90 – 265V 50/60Hz and built-in battery packs Weightaprox. 7 kg Dimensions390 x 310 x 180 mm DisplayLCD de 5,6” Transmission of measurement resultsUSB or Bluetooth

Standard accesories:

Test lead banana plug; 3 m; 10kv; red Test lead banana plug; 3 m; 10 kv; blue Test lead banana plug; 1,8 m; 10 kv; black; shielded Usb cable “Crocodile” clip 5,5 kv; black “Crocodile” clip 5,5 kv; blue “Crocodile” clip 5,5 kv; red Pin probe 10 kv with banana connector; red Pin probe 10 kv with banana connector; black Carrying case l4 for accesories Power cord Battery pack (built-in) Software Calibration certificate

AMPERIS PRODUCTS S.Lagricultura, 34, Contact 27003 Agricultura,34 +T [+34] 982 20 99 20 | F [+34] 982 20 99 11 lugo [spain]27003, Lugo, Spain info@amperis.com | www.amperis.com

Got a test to solve? Talk to an engineer.

Tell us the asset, the voltage and the test you need, and we will propose the right instrument.