Insulation Resistance Meter

HIGHEST ACCURACY & LOWEST COST

Image: Sonel insulation resistance meter in a rugged case with the lid open; panel with green display, rotary selector and terminals.

AMIC-5010

Insulation resistance measurement - 15TΩ: * 50 – 1000 V, 10 V steps * 1000 – 5000 V, 25 V steps Memory of 999 measurement results and PC transmission. Capacitance and DC/AC voltage measurement. Leakage current measurement. Digital filters function. Measurement of protective connections and equipotential bonding.

AMPERIS PRODUCTS S.Lagricultura, 34, Contact 27003 Agricultura,34 +T [+34] 982 20 99 20 | F [+34] 982 20 99 11 lugo [spain]27003, Lugo, Spain info@amperis.com | www.amperis.com

Main characteristics are the following: Insulation resistance measurement:

Insulation resistance measurement Measurement range acc. to EN 61557-2: 50 kΩ...15,0TΩ (I ISOnom =1,2 mA ó 3 mA)

  • Measurement voltage:
RangeResolutionAccuracy
0,0...999,9kΩ1kΩ
1,000...9,999MΩ0,01MΩ
10,0...99,99MΩ0,1MΩ±(3%.m.v.. + 10 digits)
100,0...999,9MΩ1MΩ±(3%.m.v.. + 10 digits)
1,000...9,999GΩ0,01GΩ
10,0...99,99GΩ0,1GΩ
100,0...999,9GΩ1GΩ±(3,5%.m.v.. + 10 digits)
1,000...9,999TΩ0,01TΩ±(7,5%.m.v.. + 10 digits)
10,0...15,00TΩ0,1TΩ±(10%.m.v.. + 10 digits)

*50 – 1000 V, 10 V steps *1000 – 5000 V, 25V steps -Continuous indication of measured insulation resistance or leakage current.

  • Automatic discharge of measured object capacitance voltage.
  • Acoustic signaling of 5 seconds intervals to facilitate capturing time characteristics.
  • Metered T1 , T2 and T3 test times for measuring one or two absorption coefficients from the range of 1…600 s.
  • Adjustable measuring time to 99’59’’. -Polarization index (PI) and dielectric absorption ratio (DAR) measurement. -Indication of actual test voltage during measurement. -1.2 mA and 3 mA test current.
  • Step voltage insulation resistance measurement (SV). -Dielectric Discharge calculation (DD). -Protection against measuring live objects. -Measurements with test leads up to 20 m. Digital filters function for measurements in high noise environment (10 s, 30 s, 60 s). Continuity measurement of protective connections and equipotential bonding in accordance with EN 61557-4 with current > 200 mA.

Values of measured resistance depending on measurement voltage

Voltage UISOMeasurement range
250V500GΩ
500V1,00TΩ
1000V2,00TΩ
2500V5,00TΩ
5000V15,00TΩ

Adjustable limits for measured resistance RISO and RCONT. Measurement of leakage current during insulation resistance testing.

Measurement of capacitance during the measurement of RISO. DC and AC voltage measurement in the range of 0…600 V. 990 cells of memory (11880 records) with the capability of wireless data transmission to a PC (with the USB-OR adapter) or through a USB cable.

Measurement of leakage current

RangeResolutionAccuracy
0…1,2 mA*Resolution and units result from the measurement range ofCalculated basing on resistance
0…3 mA*individual insulation resistance.measurements.
* - depending on the setting.

Power supply from main power line or battery packs, low battery warning indicator, built-in fast charger.

Step voltage insulation resistance measurement

Measurement of capacitance
Range ResolutionAccuracy
1...999 nF 1 nF±(5%m.v. + 5 digits)
1,00...49.99 μF 0,01 μF±(5%m.v. + 5 digits)
Capacity measurement result is displayed after the RISO measurement.
Voltaje de prueba finalSecuencia del Voltaje de prueba
1 kV200, 400, 600, 800, 1000 V
2.5 kV0.5, 1, 1.5, 2, 2.5 kV
5 kV1, 2, 3, 4, 5 kV
Duration of each “step” adjustable from 30 s to 5 mins. Measurement result for each voltage step is stored in memory.

Continuity measurement of protective connections and equipotential bonding, I=200mA

Medición de la tensión continua y alterna

Range ResolutionAccuracy
0.00...19.99 Ω 0.01 Ω±(2%m.v. + 3 digits)
20.0...199.9 Ω 0.1 Ω±(2%m.v. + 3 digits)
200...999 Ω 1 Ω±(4%m.v. + 3 digits)
Voltage on open terminals: 4...24 V. Output current at R <15 Ω: min. 200 mA (ISC : 200...250 mA). Compensation of test lead resistance. Current flowing in both directions, mean value of resistance is displayed.
RangeResolutionAccuracy
0,0...29.9 V0,1 V±(2%m.v. + 20 digits)
30,0...299.9 V0,1 V±(2%m.v. + 6 digits)
300...600 V1 V±(2%m.v. + 2 digits)
Frequency range: 45...65Hz.
“m.v.” = “measured value”.

AMPERIS PRODUCTS S.Lagricultura, 34, 27003 Agricultura,34 lugo [spain]27003, Lugo, Spain

Specifications AMIC-5010

Electric security:

Type of insulationDouble, acc. to EN 61010-1 and IEC 61557 Measurement categoryCAT IV 600V (III 1000V) EN 61010-1 Above sea level3000 m. Protection class acc. to EN 60529IP54 (IP67 with cover closed) Power supply90 – 265V 50/60Hz and built-in battery packs Weightaprox. 7 kg Dimensions390 x 310 x 170 mm DisplayLCD Measurement results memory990 cells of memory (11880 records) Transmission of measurement resultsUSB or wireless interface

Other technical data:

Standard accesories:

USB cable Test lead banana plug; 1,8 m; 10kv; red Test lead banana plug; 1,8 m; 10 kv; blue Test lead banana plug; 1,8 m; 10 kv; black; shielded “Crocodile” clip 5,5 kv; black “Crocodile” clip 5,5 kv; blue “Crocodile” clip 5,5 kv; red Pin probe 5,5 kv with banana connector; red Pin probe 5,5 kv with banana connector; black Carrying case L4 for accesories Power cord Battery pack (built-in) Software Calibration certificate

Optional accesories:

Test lead banana plug; 3 m; 10kv; red Test lead banana plug; 3 m; 10kv; blue Test lead banana plug; 3 m; 10kv; black; shielded Test lead banana plug; 5 m; 10kv; red Test lead banana plug; 5 m; 10kv; blue Test lead banana plug; 5 m; 10kv; black; shielded Test lead banana plug; 10 m; 10kv; red Test lead banana plug; 10 m; 10kv; blue Test lead banana plug; 10 m; 10kv; black; shielded Test lead banana plug; 20 m; 10kv; red Test lead banana plug; 20 m; 10kv; blue Test lead banana plug; 20 m; 10kv; black; shielded Carring backpack L-7 OR-1 radio receiver for data transmission

Image: USB dongle in a transparent housing with its removable cap.

Receiver – interface for radio transmission OR1 (USB).