Skip to content

Accessible version

RAPIDOX 5100

Accessible HTML version of the document: the same text as the PDF, with headings, tables and a description of each image. Download the original PDF (PDF · 6 pages · 207 KB)

Image: Green portable gas analyser fitted in an open black carrying case, with display and printer.

RAPIDOX 5100 MULTIGAS ANALYZER

The Rapidox 5100 Multigas Analyser is a high specification portable instrument designed for the analysis, control and monitoring of gas in a wide range of applications including; biogas, syngas, anaerobic digestion and fermentation processes.

Amperis Products, S.L. C/Barbeito María, 14 27003, Lugo, Galicia, España

Contact: Phone : +34 982 209 920

Up to six gases are simultaneously measured using a range of high precision gas sensors; each sensor is specially designed and calibrated to avoid cross interference effects with background process gas, Safety is ensured by the inclusion of flashback arrestors in the gas measuring circuit where required.

Measurable gases include oxygen (O2), carbon monoxide (CO), carbon dioxide (CO2), ozone (O3), moisture (H2O), hydrogen (H2), hydrogen sulphide (H2S), nitric oxide (NO), nitrogen dioxide (NO2), nitrous oxide (N2O), sulphur dioxide (SO2), chlorine (Cl2), methane (CH4) and ethylene (C2H4).

When configured for applications where the gases contain energy (e.g. Biogas, Syngas) the calorific value of the gas sample is determined using thermodynamic calculations and simultaneously logged and displayed on-screen.

An optional pump enables two modes of operation. For samples that are taken from a gas source at atmospheric pressure or below, the pump is activated to draw a sample through the analyser. Alternatively, the pump can be deactivated when sampling from a source at a greater atmospheric pressure, allowing the gas to flow through the analyser. Gas flow is regulated manually via a rotary knob on the fascia and displayed electronically on the screen.

Please contact Cambridge Sensotec for further information or to discuss your requirements.

Though highly configurable• Bespoke sensor combination• Continuous data logging downloaded
to suit individual customer• 7" full-colour touchscreenvia USB.
requirements, the Rapidox 5100• Lithium battery provides 8• Multi-language
range possesses a number ofhours of operation• Charges on worldwide
standard features to enhance• Heavy duty IP66 casemains voltage
functionality.• Password protection• Integrated thermal printer

Applications

Image: Biogas plant with a domed storage tank in a green meadow surrounded by trees.

Biogas

Combustion

Emissions

Gas

Research & Development

Syngas

Accesories

Image: Six numbered accessories: 1 case with gas cylinders, 2 gas panel with valves and gauges, 3 orange unit, 4 long-stem probe, 5 adapter with O-rings and filter, 6 sensor with cable.
  • 1)Calibration Kit
  • 2)Calibration Service
  • 3)Gas Recovery Bag
  • 4)Sample Probe
  • 5)Sample Probe Filters
  • 6)Collapsible Sample Probe
SPECIFICATION
Ambient Operating Temperature-10°C to 40°C
Warm-up Time3-4 minutes at 20o C
Measurement TimeApproximately 2-4 minutes (dependant on sensor configuration)
Battery LifeIn excess of 8 hours (up to 500 cycles). 4-6 hour charge
Voltage (Charging)90-260VAC, 50/60Hz
Sample Connections4mm ID/6mm OD Rectus style, closed coupled fittings
Data OutputExcel compatible data via USB memory stick
Data Storage4GB internal data storage allowing for approximately 1 year of continuous monitoring
Optional Pump0-1 litres per minute
CalibrationZero and span calibration with two user selectable gas compositions
Display7" (180mm) full-colour LCD with touchscreen operation
PrinterThermal printer allows output of results on demand
Analyser Dimensions180mm(H) x 480mm(W) x 360mm(D)
Weight7kg
Image: Rapidox portable gas analyser in an open rugged case, green front panel with built-in printer and screen listing gas readings such as CH4, CO2, CO, H2 and energy.

Rapidox 5100 Sensor Matrix

GasO2CO2CH4 CH4H2OCOCOCO (H2 resist)Cl2NONO2C2H4O3H2SHe/H2NH3SO2 SO2
Sensor TypeEC-EIRIR LTDCAPIRECECECEECECIRECECTCDECEC IR
ife span (mth)60>60>60 >60>36>602424121212>601224>602424 >60
Cal. (mth)121212 121212121266612612121212 12
0-100%
0-80%
0-60%
0-50%
0-30%
0-20%
0-10%
0-5%
0-3%
0-2%
0-1%
0-5,000ppm
0-3,000ppm
0-2,000ppm
0-1,000ppm
0-500ppm
0-250ppm
0-200ppm
0-100ppm
0-60ppm
0-20ppm
0-10ppm
-65oC to+20oC
-100oC
0 to 5 bar
0 to 10 bar
-1 to 0 bar

Got a test to solve? Talk to an engineer.

Tell us the asset, the voltage and the test you need, and we will propose the right instrument.

+34 982 20 99 20 Mon–Thu 09:00–14:00 · 15:30–19:00Fri 08:00–14:00 Request a quote