Rapidox 2100-FGA: gas analyser

SF6 gas analysers: purity, moisture, SO2See the comparison

Rapidox 2100-FGA: gas analyser

O2 range
10E-30 ppm to 30 %
Response
≈4 s for a 90 % step change
Supply
90-260 V AC, 50/60 Hz
Dimensions and weight
250 × 263 × 150 mm
Working pressure
Up to 10 bar (200 bar burst)
Operating conditions
5 °C to 35 °C

Which model do you need? Talk to an engineer

+34 982 20 99 20

ISO 9001 · ISO 14001 · ISO 45001

Forming gas analyser with extended ultra-low oxygen range

Controlling a forming gas atmosphere —nitrogen with hydrogen— means measuring extremely low oxygen partial pressures and knowing its moisture content at the same time. The Rapidox 2100-FGA extends its zirconia sensor range down to 10E-30 ppm O2, with ±1 % of the actual concentration or 0.5 ppm accuracy and a typical response under 4 s, and calculates the H2O dewpoint thermodynamically from the hydrogen content entered by the operator; with insertion probes it works in gas up to 1000 °C.

Rapidox 2100-FGA key data

  • Oxygen measurement: extended range of 10E-30 ppm to 30 % O2, accuracy of ±1 % of the actual concentration or 0.5 ppm (whichever is greater) and a typical response under 4 s for 90 % of a step change.
  • Moisture without an extra sensor: H2O dewpoint calculated thermodynamically from the hydrogen content, shown in °Cdp or ppmV alongside oxygen; a pure hydrogen mode is included.
  • Temperature: gas up to 650 °C, or up to 1000 °C with insertion probes; temperature is read at the sensor head up to 660 °C and above that by the standard type K thermocouple (0-1250 °C, ±1 °C).
  • Pressure and sample: working pressure up to 10 bar (200 bar burst), vacuum-tight below 10E-4 Torr and a flow rate of 0.1 to 4 L/min (1 L/min recommended), or static gas.
  • Sensor: pre-calibrated hot-swap design, replaced without a PC; life above 17,500 h and a 2 m high-temperature cable extendable to 25 m.
  • Auxiliary measurements: socket for standard 4-20 mA transmitters, configurable for pressure or vacuum; optional pressure sensor for −1 to 0 bar, 0-5 and 0-10 bar.
  • Alarms and outputs: programmable high and low alarm relays, 0-5 V (min. 5 kΩ load) and 4-20 mA (max. 500 Ω load), RS232 (RS485 optional) and Modbus RTU/Ethernet, 20 × 4 character OLED display and pin-code protection.
  • Installation: 19″ panel mount (300 mm × 4U) or IP65 wall housing, 250 × 263 × 150 mm and 3.5 kg with sensor, 90-260 V AC with 30 W maximum; operation from 5 °C to 35 °C, 900-1,100 mbar absolute and 10-90 % RH.

Advantages

A range built for forming gas

The zirconia sensor extends its measurement down to 10E-30 ppm O2, covering the ultra-low partial pressures of forming gas mixtures; with insertion sample tubes it operates up to 1000 °C.

Dewpoint without a moisture sensor

The analyser resolves thermodynamic equations to deliver the H2O dewpoint of the mixture: the operator enters the hydrogen content and the value is displayed in °Cdp or ppmV alongside the oxygen reading.

Fast, continuous response

Analysis is on-line, with about 4 s for a 90 % response to a step change; temperature is read at the sensor head up to 660 °C and by the standard type K thermocouple above that.

Hot-swap sensor

The pre-calibrated sensor is replaced on site without a PC, keeping downtime to a minimum.

Connected to the process

Programmable alarm relays and analogue outputs, RS232 and Modbus RTU communications, auxiliary socket for 4-20 mA transmitters and data-logging software.

Applications

  • Hydrogen forming gas in metal heat treatment.
  • Annealing, sintering and brazing atmospheres.
  • Manufacturing under protective atmospheres.
  • Research and development.

About the product

The Rapidox 2100-FGA is a special variant of the zirconia oxygen analyser for continuous duty, panel mounted (19-inch) or in a wall-mountable IP65 housing, with sensors up to 25 m away. It complies with EMC Directive 2004/108/EC and carries UL/ETL certification UL-61010-1.

Available accessories: calibration kit, calibration service, multiplex sampling system, sample pump, wall-mount cabinet, auxiliary pressure sensor, vacuum fittings (CF50, ISO KF25), stainless probe, filters and thermal printer.

Technical specifications

O2 range10E-30 ppm to 30 %; ±1 % of reading or 0.5 ppm
Response≈4 s for a 90 % step change; sensor life over 17 500 h
H2O readingDewpoint by thermodynamic calculation, °Cdp or ppmV
Maximum gas temperature650 °C (1000 °C with insertion probes)
ThermocoupleType K, 0-1250 °C, ±1 °C (included)
Outputs0-5 V, 4-20 mA, RS232/RS485, Modbus RTU/Ethernet
Supply90-260 V AC, 50/60 Hz
Dimensions and weight250 × 263 × 150 mm; 3.5 kg with sensor
Working pressureUp to 10 bar (200 bar burst); vacuum-tight below 10E-4 Torr
Sample flow0.1-4 L/min (1 L/min recommended); static gas or vacuum
Operating conditions5 °C to 35 °C; 900-1,100 mbar absolute; 10-90 % RH

Frequently asked questions

Can it measure at furnace temperature?

Yes. The Rapidox 2100-FGA accepts gas up to 650 °C and reaches 1000 °C with insertion probes. Temperature is read at the sensor head up to 660 °C and above that by the standard type K thermocouple, which covers 0 to 1250 °C with ±1 °C.

Is it also suitable for pure hydrogen?

Yes. Besides the calculation for forming gas mixtures, in which the operator enters the hydrogen content, the Rapidox 2100-FGA offers a pure hydrogen mode. In both cases it measures oxygen from 10E-30 ppm with ±1 % of the actual concentration or 0.5 ppm, whichever is greater.

What is forming gas?

A nitrogen-hydrogen mixture used as a protective atmosphere in metal heat treatment; its oxygen partial pressures are extremely low, which demands the extended measurement range.

How is the dewpoint obtained?

By thermodynamic calculation: the operator enters the hydrogen content of the mixture and the analyser shows the dewpoint in °Cdp or ppmV together with the oxygen reading.

Can the sensor be replaced on site?

Yes. The pre-calibrated hot-swap sensor is exchanged locally without a PC.

Which mounting options exist?

Panel mount (19-inch) or wall-mountable IP65 housing, with the sensor remote up to 25 m.

Videos

Rapidox 2100-FGA (4:31)
Rapidox 2100-FGA: forming gas analyser that measures oxygen with a zirconia sensor, down to 10E-30 ppm, and calculates the dewpoint. What forming gas is, why its moisture is read through oxygen, and how to install the sensor, calibrate it and read it. Watch the video with its transcript

AI-generated video. The information must be validated by an Amperis engineer.

Downloads

PDFDownload the catalogue · Rapidox 2100-FGA: gas analyser207 KB · EN Accessible version (HTML)

More in SF6 gas analysers: purity, moisture, SO2

See all →

Got a test to solve? Talk to an engineer.

Tell us the asset, the voltage and the test you need, and we will propose the right instrument.

+34 982 20 99 20 Mon–Thu 09:00–14:00 · 15:30–19:00Fri 08:00–14:00 Request a quote